  4. Validation of Multi-Frequency Bioelectrical Impedance Analysis Versus Dual-Energy X-Ray Absorptiometry to Measure Body Fat Percentage in Overweight/Obese Colombian Adults
Details

Validation of Multi-Frequency Bioelectrical Impedance Analysis Versus Dual-Energy X-Ray Absorptiometry to Measure Body Fat Percentage in Overweight/Obese Colombian Adults

Journal
American Journal of Human Biology
ISSN
1042-0533
Date Issued
2018
Author(s)
Garcia-Hermoso, A  
Abstract
Objectives: To verify the validity of multi-frequency bioelectrical impedance analysis (mBCA) for predicting body fat percentage (BF%) in overweight/obese adults using dual-energy X-ray absorptiometry (DXA) as the reference method. Methods: Forty-eight adults participated (54% women, mean age = 41.0 ± 7.3 years old). The Pearson s correlation coefficient was used to evaluate the correlation between BIA and BF% assessed by DXA. The concordance between BF% measured by both methods was obtained with Lin s concordance correlation coefficient and Bland–Altman difference plots. Results: Measures of BF% were estimated as 39.0 (SD = 6.1) and 38.3 (SD = 6.5) using DXA and mBCA, respectively. The Pearson s correlation coefficient reflected a strong correlation (r =.921, P =.001). The paired t-test showed a significant mean difference between these methods for obese men BF% of −0.6 [(SD 1.95; 95% CI = −4.0 to 3.0), P =.037]. Overall, the bias of the mBCA was −0.6 [(SD 2.2; 95% CI = −5.0 to 3.7), P =.041], which indicated that the mBCA method significantly underestimated BF% in comparison to the reference method. Finally, in both genders, Lin s concordance correlation coefficient showed a strong agreement. More specifically the DXA value was ρc = 0.943 (95% CI = 0.775 to 0.950) and the mBCA value was ρc = 0.948 (95% CI = 0.778 to 0.978). Conclusions: Our analysis showed a strong agreement between the two methods as reflected in the range of BF%. These results show that mBCA and DXA are comparable methods for measuring body composition with higher body fat percentages. However, due to broad limits of agreement, we can only recommend mBCA for groups of populations. © 2017 Wiley Periodicals, Inc.
Get Involved!
  • Source Code
  • Documentation
  • Slack Channel
Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your Institution's web identity.

Need professional help?

The original creators of DSpace-CRIS at 4Science can take your project to the next level, get in touch!

Logo USACH

Universidad de Santiago de Chile
Avenida Libertador Bernardo O'Higgins nº 3363. Estación Central. Santiago Chile.
ciencia.abierta@usach.cl © 2023
The DSpace CRIS Project - Modificado por VRIIC USACH.

  • Accessibility settings
  • Privacy policy
  • End User Agreement
  • Send Feedback
Logo DSpace-CRIS
Repository logo COAR Notify